Product Details:
|
Type: | Patient Monitor Machine | Standard Parameters: | ECG, RESP, SpO2, PR, NIBP, Dual-channel TEMP |
---|---|---|---|
Dimension: | 314 Mm(L) X 145 Mm(W) X 264 Mm(H) | Weight: | <3.9 Kg(excluding Battery And Accessories) |
Display: | 12.1" TFT Color LCD | Interface: | Multi-language |
High Light: | hospital vital sign machine,vital sign monitors |
ICU Multiparameter Patient Monitor Machine / Vital Sign Monitors
FL-PM8000 Introduction
The monitor has abundant functions that can be used for clinical monitoring with adult, pediatric and neonate. Users may select different parameter configuration according to different requirements. The monitor, power supplied by 100-240V~,50/60Hz, adopt 12.1” color TFT LCD displaying real-time date and waveform. It can synchronously display eight-channel waveform and full monitoring parameters equipped with an optional 48mm thermal recorder. The monitor can be connected to the central monitoring system via wire or wireless network to form a network monitoring system.
This device can monitor such parameters as ECG, RESP, NIBP. SpO2, and dual-channel TEMP, ect. It integrates parameter measurement module, display and recorder in one device to form a compact and portable equipment. Its replaceable internal battery brings lot of convenient for patient moving.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
ECG
Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis
RESP
Respiration rate(RR)
Respiration waveform
SpO2
Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
NIBP
Systolic pressure(SYS), Diastolic pressure(DIA), Mean
pressure(MEAN)
TEMP
TI, T2, TD
IBP(optional)
CH1: SYS, DIA, MAP
CH2: SYS, DIA, MAP
IBP waveform
CO2(optional)
EtCO2
InsCO2: Inspired Minimum CO2
AWRR: Air Way Respiration
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
Accessories
Adult fingertip SpO2 probe (5-pin)
Adult NIBP cuff
NIBP extension tube
ECG lead cable
ECG electrode
Temperature probe
Power cord
User Manual
Physical characteristic
Dimension: 314 mm(L) x 145 mm(W) x 264 mm(H)
Weight: <3.9 kg(excluding battery and accessories)
Features
l 12.1" TFT color LCD, multi-language interface (English, French, German, Italian, Dutch, Russian, Portuguese, Turkish, Spanish(EU),Polish, Roumania, Kazakh, Czech, Trad Chinese, Spanish(MX), Bulgarian, Chinese, Ukraine).
l Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
l All-round monitor for adult, pediatric and neonate.
l With standard interface, oxygen graph, trend graph, big character interface and view bed, convenient to observe.
l Finish all operations by keys and knobs.
l Maximum 8-channel waveform synchronous display.
l Display 7-lead ECG waveform on one screen, cascade ECG waveform display.
l Adopt digital Sp02 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
l Heart rate variability (HRV) analysis function.
l NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
l Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
l Drug concentration calculation and titration table functions.
l One-touch printing of trend graph.
l Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
l AC/DC, built-in rechargeable lithium battery achieve uninterrupted
monitoring.
l Anti-high frequency surgical unit, defibrillation-proof (special leads are necessary).
Performance
ECG
Lead mode: 3-lead: 1, II, II 5-lead: I, II, II, aVR, aVL, aVF, V
Waveform: 3-lead, 1-channel 5-lead, 2-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV
Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR: Measurement and alarm range: 15-350 bpm
Resolution: 1 bpm
Accuracy: +1% or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0 mV ~+2.0 mV
Accuracy: -0.8 mV~ +0.8 mV: ±0.04 mV or ± 10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy:±2 rpm
Apnea alarm: 10 - 40 s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 - 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: + 3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
NIBP
Over-pressure protection:
Adult mode: 297 mmHg ± 3 mmHg
Pediatric mode: 240 mmHg ± 3 mmHg
Neonatal mode: 147 mmHg ± 3 mmHg
SpO2
Measurement and alarm range: 0~100 %
Resolution: 1%
Accuracy:70~100 %: ±2 %; 0 ~ 69 %: unspecified
PR
Measurement and alarm range:25~250 bpm
Measurement accuracy: ±2 bpm or ± 2 %, whichever is greater
TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ± 0.1 ℃
EtCO2
Method: Sidestream or Mainstream
Measuring Range: 0~150 mmHg
Resolution: 0 ~ 69 mmHg: 0.1 mmHg, 70 - 150 mmHg: 0.25 mmHg
Accuracy: 0 - 40 mm Hg: ±2 mm Hg,41 - 70 mm Hg: ± 5 %, 71 - 100 mm Hg: ± 8 %, 101 - 150 mm Hg: ± 10 %
AWRR Range: 2~150 rpm
AWRR Accuracy: ±1rpm
Apnea Alarm: Yes
IBP
Channel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, PI, P2
Measuring and Alarm Range: -10 ~ 300 mmHg
Resolution: 1 mmHg
Accuracy: ± 2% or 1mmHg, whichever is greater
Power supply: 100-240V, 50/60 Hz
Safety classification: Class I, type CF defibrillation-proof applied part
Contact Person: Amanda
Tel: 15136250979